Monday, August 22, 2011

Deuterium Lamp

It is known within the industry that deuterium lamps exhibit reduced intensity as they age during their normal service life.
Lamp Operating Specifications
Detector Type:
Dual Wavelength Absorbance
Light Source:
Deuterium Lamp
Power Consumption:
Approx. 30 Watts
Drift:
±0.3% per hour Max.
Operational Anode Voltage:
85 VDC (Typical)
Operational Anode Current:
300 ±30mADC
Discharge Trigger Voltage:
More Than 400VDC
Preheat Voltage:
3.0V ±.3 V, AC or DC
Preheat Current:
Approx. 5 Amps
Preheat Time:
20 Sec. Min.
Operating Voltage:
0 to 1 V
Operating Current:
0 to1.8 A
Lifetime:
2000 hours